| LED display full English, English menu operation |
| Multi-parameter display in the screen: It displays several parameters in the same screen, including turbidity, |
| output current, time and status |
| History curve: The secondary meter can save the measurement data automatically every 5 minutes, and can store |
| the turbidity data of one consecutive month. |
| Intelligent: ZDYG-2088Y/T industrial turbidity meter uses high precision AD conversion and single chip range |
| micro-processing technology, and has such functions as turbidity measurement, automatic switch of measurement |
| and self-check of the instrument. |
| Strong interference immunity : Photoelectric coupling separation technology is used for the current output, |
| providing high resistance against interference and allowing for long transmission. It also has high electromagnetic |
| compatibility. |
| Waterproof and dustproof design: The protection grade is IP65, and it is suitable for outdoor use. |
| RS485 communication interface: It can be easily connected to a computer to perform monitoring and |
| communication |
| 1.Measuring range:0~10NTU,0~20NTU,0~100NTU,0~500NTU,0~1000NTU |
| 0~2000NTU,0~4000NTU( specified when ordering ) |
| 2. Resolution:0.001NTU,0.01NTU,0.1NTU,1NTU |
| 3. Accuracy: ± 2 % |
| 4. Isolated current output: 4~20 mA(load <750 Ω ) |
| 5. Alarm relay: AC 220V, 7 A |
| 6. Communication interface: RS485(optional) |
| 7 . Power supply:AC220 ± 22V,50 ± 1Hz,DC24V ± 2.4(optional) |
| 8. Protection grade: IP65 |
| 9. Data storage:1 month(1point/5minute) |
| 10. Power-down memory: 10 years |
| 11. Weight: 0.8kg |
| 12. Overall dimension: 146 (length)x 146 (width)×1 08 (depth)mm |
| Dimension of the hole: 138 x 138 mm |
| 13. Working conditions: ambient temperature: 0-60 ºC ; relative humidity <85 % |
| 14.Installation type of sensor: Flow-through or immerged (specified when ordering) |